Disinfection and pre-sterilization cleaning of surgical instruments and reusable medical devices
Патрабаваньні выкананьня
Authors: Rednenko V.V., Brikez J.I.
Editor: Rednenko V.V.
Equipment for the practical skill:
- disinfection table
- sink for processing instruments
- distilled water in a container
- containers with a solution of a chemical disinfectant with detergent properties:
- "Disinfection of MI"
- "Disinfection of dressings, PPE"
- "Washing of MI"
- tweezers in a container with a working solution of a chemical disinfectant
- protective gloves (not sterile), mask, glasses
- brush, cotton-gauze swab
- surgical instruments: tweezers, scissors, clamp, needle holder
- gloves, mask, glasses
Clinical situation:
- Disinfect and pre-sterilize surgical instruments after surgery. Pre-sterilization cleaning is combined with disinfection in one stage.
Order of the skill performance:
Preparatory stage:
- wash your hands (см. Hygienic hand washing)
- put on mask, glasses, gloves
Main stage:
- clean the MI from visual contaminations
- open the container "Washing of MI", place the container lid near the container
- immerse all MI in this container
- open MI with locking parts, made several working movements with them in the solution
- MI with channels, flush with disinfectant solution with a syringe
- disassemble MI, consisting of several parts
- if necessary, clean the MI from visual contamination manually using a cotton-gauze swab, place the used cotton-gauze swab into a container "Disinfection of dressing material, PPE"
- disinfect MI
- open the container "Disinfection of MI", place the container lid near the container, place the press plate inside the container vertically beside one of the walls
- place the washed MI into the container
- press the MI with the press plate (layer of solution must be at least 1 cm above the plate) and close the container with a lid
- withstand the exposure specified in the recommendations for the use of this disinfectant (say loudly *** minutes have passed)
- open the container "Disinfection of MI", place the container lid on the container or next to it, put the press plate inside the container vertically against one of the walls
- wash each instrument in the same solution and container
- open MI with locking parts, made several working movements with them in the solution
- MI with channels, flush with disinfectant solution with a syringe
- disassemble MI, consisting of several parts
- use a gauze for washing
- use a brush to wash cavities and channels (except for rubber products)
- remove the washed MI from the container (one at a time) and place them on a perforated tray (made of heat-resistant material)
- rinse MI
- place the perforated tray with MI in the sink and rinse them under running water
- metal MI for 3 minutes
- polymer MI for 10 minutes
- rinse the instruments on the perforated tray with distilled water for 30 seconds
- place the perforated tray with MI in the sink and rinse them under running water
- dry MI
- place the perforated tray from the sink on a metal non-perforated tray (MI with locking parts, must be opened, collapsible MI, disassembled into component parts)
- place the tray with MI in a dry oven
- dry all instruments with hot air at a temperature of 85 * C until the moisture completely disappears (say loudly "Medical items are dried")
- wait for the instruments to cool down (say loudly "Instruments are cold"
- check the quality of pre-sterilization cleaning
- take a gauze pad and moisten it with azopyram working solution
- wipe with this pad 1% of the total number of MI (but not less than 3 pcs.)
- evaluate the color change of the pad and draw a conclusion about the quality of the pre-sterilization treatment
Final stage:
- remove gloves and place them in the container “Disinfection of dressings, PPE” (see practical skill«Removal contaminated gloves»);
- take off a mask and place it in the container “Disinfection of dressings, PPE”;
- take off glasses;
- wash your hands (см. Hygienic hand washing)